Description
Fluke DSX CERTIFIBER Calibration
The Fluke DSX CERTIFIBER is a fiber optic test set, where the calibration has to cover the optical side as well as the electronics. Power measurement accuracy is wavelength-dependent, so a meter correct at 1310 nm can be out at 1550 nm, and a single-wavelength check would never reveal it.
Calibration of the Fluke DSX CERTIFIBER is performed at June Company for fiber optic installers, network engineers, data centre technicians, and telecommunications crews. Every point is measured against reference standards, with as-found readings recorded before any adjustment and as-left readings afterwards, so you can see what condition the instrument was actually in when it arrived.
What we verify on the Fluke DSX CERTIFIBER
- Optical power measurement at each supported wavelength – accuracy is wavelength-specific and does not transfer between them.
- Source output power and stability.
- Loss measurement and referencing.
- OTDR distance and event-location accuracy where fitted.
- Optical connector endface condition.
- Adapter and launch cord condition.
FAQs
How often should this be calibrated? Most quality systems work on an annual interval, though how hard the instrument is used and how critical its readings are should drive that decision rather than the calendar alone.
Do I get as-found data, or only the final result? Both. As-found readings are recorded before any adjustment and as-left afterwards, which is what tells you whether the instrument was in tolerance during the work you already did with it.
Calibration is documented and returned with the instrument. If we find a fault that calibration cannot correct, we quote the repair before going any further. Email us or submit a service request.



Reviews