Description
Fluke 754 Calibration
The Fluke 754 is a process calibrator, which is unusual to calibrate because it both sources and measures. Each direction has to be verified independently – an instrument can source 12.000 mA accurately while reading an applied 12.000 mA incorrectly, and only checking one direction hides that.
The Fluke 754 is calibrated at June Company for process instrument technicians, plant calibration teams, and field service engineers. Every point is measured against reference standards, with as-found readings recorded before any adjustment and as-left readings afterwards, so you can see what condition the instrument was actually in when it arrived.
What we verify on the Fluke 754
- 24V loop supply output.
- mA source and mA measure, verified separately in each direction.
- Voltage source and measure across ranges.
- Each RTD and thermocouple type supported.
- Frequency and resistance functions.
- Pressure measurement through each module used.
FAQs
Can it be adjusted if it has drifted? Where the instrument supports adjustment, yes – and the as-found and as-left readings are both recorded so the change is visible.
Do I get as-found data, or only the final result? Both. As-found readings are recorded before any adjustment and as-left afterwards, which is what tells you whether the instrument was in tolerance during the work you already did with it.
Calibration is documented and returned with the instrument. If we find a fault that calibration cannot correct, we quote the repair before going any further. Email us or submit a service request.