Fluke TiR32 Thermal Imager Repair

$245.00

🛡️ 90-Day Warranty ✓ Fee Credited to Repair ⚡ No Forms Needed

Expert Fluke Infrared Camera TiR32 repair and calibration with documented test results. Our repair process covers diagnostics through final thermal sensitivity verification.

📋 How It Works

  1. Submit Request – Pay evaluation fee & provide equipment details
  2. Ship to Us – Send your equipment to our Oregon facility
  3. We Evaluate – Detailed diagnosis & repair estimate
  4. Fee Credited – Your evaluation fee applies to repair cost
  5. We Repair – Professional repair with 90-day warranty
  • 🔧 Fast Turnaround: Most repairs are completed within 1–3 business days.
  • 📏 Precision Calibration: NIST-traceable calibrations included for most repairs.
  • 🛡️ Warranty: 90-day warranty on all repairs.
  • 📝 Transparency: Detailed evaluation and repair estimate before work begins.
✓ Evaluation Fee Credited Toward Repair

This evaluation fee is applied as credit toward your final repair cost. After we evaluate your equipment, you'll receive a detailed estimate before any repair work begins.

Description

Fluke TiR32 Thermal Imager Repair

The Fluke TiR32 is a building-diagnostics thermal imager. The TiR line was tuned for thermal sensitivity rather than raw temperature range, because finding moisture intrusion, missing insulation and air leakage means resolving fractions of a degree across a wall, not measuring a hot bearing.

June Company’s certified technicians repair the Fluke TiR32 for building inspectors, energy auditors, restoration contractors, and home-performance professionals. Every imager is repaired at component level and its temperature accuracy verified against reference standards before it ships back, with the measured results documented.

Common Fluke TiR32 problems we fix

  • Temperature readings drifted – the measurement path rather than the image – a picture that looks fine can still be reporting the wrong number.
  • Display or touchscreen failure.
  • Battery will not hold charge or unit will not power on – the pack, the charging circuit and the power section, tested separately.
  • Images not saving or SD card not recognised.
  • Loss of thermal sensitivity – the fault that matters most on this series – small temperature differences stop resolving and moisture patterns disappear into the background.
  • Image degrading between shutter cycles – the internal NUC shutter, which recalibrates the detector every few minutes – when it sticks the image drifts and blotches until the unit is power-cycled.
  • Dead pixels or fixed pattern noise – microbolometer detector faults, which show as permanent spots or a grid pattern that no amount of re-focusing clears.
  • Humidity or dew-point calculation errors.

FAQs

What if it is not worth repairing? We say so in the quote. The evaluation is free and there is no obligation to proceed.

How long does a repair take? Most are completed in 5-10 business days, with a firm quote after evaluation and before any work starts.

Every repair carries a 90-day parts-and-labor warranty, with a firm quote before any work begins. Email us or submit a service request.

Brand

Fluke Repair

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.