Keysight MSOX6004A Oscilloscope Repair

$245.00 evaluation fee

🛡️ 90-Day Warranty ✓ Fee Credited to Repair ⚡ No Forms Needed

📋 How It Works

  1. Submit Request – Pay evaluation fee & provide equipment details
  2. Ship to Us – Send your equipment to our Oregon facility
  3. We Evaluate – Detailed diagnosis & repair estimate
  4. Fee Credited – Your evaluation fee applies to repair cost
  5. We Repair – Professional repair with 90-day warranty
  • 🔧 Fast Turnaround: Most repairs are completed within 1–3 business days.
  • 📏 Precision Calibration: NIST-traceable calibrations included for most repairs.
  • 🛡️ Warranty: 90-day warranty on all repairs.
  • 📝 Transparency: Detailed evaluation and repair estimate before work begins.
✓ Evaluation Fee Credited Toward Repair

This evaluation fee is applied as credit toward your final repair cost. After we evaluate your equipment, you'll receive a detailed estimate before any repair work begins.

Description

Keysight MSOX6004A Oscilloscope Repair

The Keysight MSOX6004A Oscilloscope is a bench oscilloscope. Its accuracy depends on the analog front-end ahead of the digitizer – attenuators, relays and amplifiers that are switched every time a technician changes a range, and which therefore wear in a way the digital section never does.

June Company repairs and calibrates the Keysight MSOX6004A Oscilloscope for electronics engineers, R&D laboratories, service departments, and production test teams. Every instrument is repaired at component level and verified against reference standards before it ships back, with the measured results documented.

Common Keysight MSOX6004A Oscilloscope problems we fix

  • Internal power supply faults.
  • GPIB, LAN, USB or remote-control faults – which breaks automated test sequences while the instrument still works from the front panel.
  • Channel dead, noisy, or inaccurate – input front-end damage, usually from an over-range input, repaired at component level.
  • Attenuator relays intermittent – these click on every range change and eventually stop making reliable contact, which shows as an amplitude error on some ranges and not others.
  • Vertical amplitude accuracy drifted.
  • Timebase accuracy out of specification.
  • Trigger unstable or not arming.
  • Display, backlight or touchscreen faults.
  • Cooling fan or thermal shutdown.

FAQs

One channel is dead. Worth repairing? Usually yes, and it is a contained repair. Input front-ends fail individually.

What if it is not economical to repair? We tell you in the quote. The evaluation is free and there is no obligation to proceed.

Every repair carries a 90-day parts-and-labor warranty, with a firm quote before any work begins. Email us or submit a service request.

Related Keysight repairs

We service 319 Keysight instruments. See all Oscilloscope Repair.

Brand

Keysight Repair

Shipping & Returns

Packing your equipment. You are responsible for packaging equipment adequately for transit. We strongly recommend the original manufacturer’s case, or double-boxing with at least two inches of shock-absorbing material on all six sides. We return your equipment in the packaging it arrived in, so please send it in packaging that will survive the return trip as well. If your packaging is damaged or inadequate for the return, we will substitute suitable packaging and may charge the cost of materials.

Insurance. Equipment is not insured by June Company while in transit. We strongly recommend insuring your shipment for its full replacement value.

Risk of loss. Risk of loss remains with you until your equipment is received at our facility, and transfers back to you when your repaired unit is delivered to the carrier (F.O.B. shipping point). June Company is not responsible for equipment lost, delayed, misdelivered, or damaged in transit, in either direction.

Return label. A prepaid and preprinted return shipping label should be included with your shipment. If one is not included, we will provide one.

Transit damage. Damage occurring in transit is the responsibility of the carrier. It is not a warranty defect and is not covered by the 90-day repair warranty. Any claim must be filed by you with the carrier; we will supply photographs and documentation to support it at no charge.

Full details: Terms & Conditions, Section 6 · Return Policy

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.