Description
Hipotronics 100HVT Hipot Tester Repair
June Company Industrial Services offers expert repair and NIST-traceable calibration services for the Hipotronics 100HVT AC Hipot Tester. Our technicians specialize in high-voltage dielectric test equipment and use OEM parts to restore full system functionality. Each unit undergoes comprehensive diagnostics, component-level repairs, and rigorous performance testing.
Common Issues We Repair
-
Output voltage instability or failure to reach rated 100 kV
-
Burned or shorted internal HV transformer windings
-
Faulty or drifting voltage divider circuits
-
Arcing or insulation breakdown during testing
-
Cracked or carbon-tracked epoxy encapsulant
-
Non-functional analog metering or display drift
-
Power-up failure or blown input fuses
-
Overload protection tripping under light load
Our Repair Process
-
Comprehensive functional inspection and fault diagnosis
-
Replacement of damaged HV components with OEM parts
-
Rebuild of internal insulation where degradation is detected
-
Recalibration of voltage metering circuits
-
High-voltage output testing to full rated kV under load
-
Final performance verification and burn-in test
NIST-Traceable Calibration
After repair, we calibrate each 100HVT unit using certified standards to ensure voltage accuracy and measurement reliability. Calibration results are documented and traceable to NIST for compliance with ISO/IEC 17025 standards.
Why Choose June Company?
-
OEM-grade repairs backed by over 20 years of experience
-
Industry-leading turnaround times
-
Detailed service reports and calibration certificates included
-
Trusted by utilities, OEMs, and high-voltage labs worldwide
Click Request Repair to begin your Hipotronics 100HVT Hipot Tester repair. We’ll promptly provide pricing, turnaround times, and shipping instructions to get your equipment back in optimal condition.